Santec PEM-340 偏振消光比测试仪(PER) 可以实现快速,精准的光器件偏振轴对准及保偏光纤(PMF)性能评价。除此之外,使用应答时间为1kHz的模拟功率计, 可对应半导体激光及光纤的高速自动对准。
| 种类 |
单位
|
正常功率
|
高功率
|
|||
| 波长范围 |
nm
|
1260 to 1630
|
||||
| PER 动态范围 |
dB
|
50
|
40
|
|||
| PER 范围 |
dB
|
0 to 50
|
- 5 to + 10 dBm
|
–
|
–
|
|
|
0 to 40
|
- 15 to + 10 dBm
|
0 to 40
|
- 5 to + 20 dBm
|
|||
|
0 to 30
|
- 25 to + 10 dBm
|
0 to 30
|
- 15 to + 20 dBm
|
|||
|
0 to 20
|
- 35 to + 10 dBm
|
0 to 20
|
- 25 to + 20 dBm
|
|||
| PER 分辨率 |
dB
|
0.01
|
||||
| PER 精度 *1 |
PER < 30 dB |
dB
|
<± 0.3
|
|||
| PER ≧ 30 dB |
<± 1.0
|
<± 1.0
|
||||
| 输入功率范围 |
dBm
|
- 40 to + 10
|
- 25 to + 20
|
|||
| 功率分辨率 |
dB
|
0.01
|
||||
| 功率精度 *2 |
dB
|
<± 0.3
|
||||
| 偏振角度分辨率 *3 |
deg
|
0.12
|
||||
| 偏振角度精度 |
deg
|
<± 0.45
|
||||
| 测量速度 |
sec
|
0.1 / 0.2 / 0.4
|
||||
| 外部接口 |
–
|
GP-IB, RS-232C
|
||||
| 操作温度 |
℃
|
0 to 40
|
||||
| 体积 (w) x (D) x (H) |
mm
|
210 x 300 x 80
|
||||
| 重 |
kg
|
3
|
||||
| 工作电压 |
V
|
AC 100 to 240
|
||||
| 最大功耗 |
W
|
35
|
||||
*1: 温度: 20 - 30 ℃, 测量速度: 2.5 Hz
*2: 校准波长 : 1310 nm / 1550 nm
*3: 测量速度: 2.5 Hz
*4: SPC = 超级紧密接触, APC = 斜8度角端面